MAINTENANCE
& CALIBRATION PROCEDURE: YSI-85
MULTIMETER
(To be performed
by program staff or people who have been trained by them.)
Last revised
EQUIPMENT
NEEDED:
·
“Maintenance
& Calibration” binder
·
YSI-85
multimeters
·
YSI
5906 Membrane Cap Kit containing:
·
new
black plastic DO membranes
·
small
nylon scrub brush
·
quarter-sized
sanding disk
·
oxygen
probe electrolyte solution
·
distilled
water
·
foaming
bathroom cleaner
·
[bucket]
·
[commercial
ammonia]
·
[bleach]
·
[conductivity
calibration solution]
The
YSI-85 multimeter is a hand-held field meter that measures oxygen,
conductivity, salinity, and temperature.
As of summer 2002, Streamkeepers does not monitor salinity except in
some cases at grab-sample sites to test for tidal influence.
MAINTENANCE & CALIBRATION
NOTEBOOK:
Record
all calibration and maintenance activity in the “Streamkeepers Maintenance
& Calibration” binder kept in the Streamkeepers office. See the binder for details.
CALIBRATION OF THE YSI-85:
Calibration
requirements of the YSI-85 are as follows:
Temperature: No calibration required.
Salinity: No calibration required.
Oxygen: Calibration is performed by the monitoring
team on the day of monitoring. (See
“Water Chemistry” protocol.)
Conductivity: “System calibration is rarely required because
of the factory calibration of the YSI Model 85.
However, from time to time it is wise to check the system calibration
and make adjustments when necessary” (YSI Model 85 Operations Manual, November 1998). In practice, Streamkeepers will not recalibrate
for conductivity unless:
the cell constant is not 5.0/cm ±
4% (per the Operations Manual); or
the meter does not read between 975-1025 when inserted in a 1000 mS
standard solution (±2%, per the Operations
Manual); or
monitoring teams note erratic readings, routine maintenance (below) does
not resolve the problem, and the meters from two field kits do not agree within
4% when tested together.
If calibration is to be performed, procedure will
follow the product Operations Manual referenced above, unless a revised manual supersedes it.
MAINTENANCE OF THE YSI-85:
Just before the
quarterly monitoring period begins:
1) Clean the DO probe:
a) Unscrew the sensor guard at the end of the
probe.
b) Unscrew and discard the old DO membrane
cap.
c)
Squirt
the probe with distilled water.
d)
Gently
wet-sand the silver (Ag) anode sides and
gold (Au) cathode top, preferably with two different marked sanding disks
(provided with the Membrane Cap Kit), to avoid intermingling of the
metals. Stop when darkening deposits
have been wiped off—don’t overdo it!
Rinse with distilled water and pat dry.
e) If the probe is still stubbornly dirty and
unshiny, soak it in commercial ammonia (3% ammonium hydroxide – PAR ammonia is
fine) overnight with the meter off, the membrane removed, and the membrane
solution thoroughly rinsed off. The next
day, rinse under running water for 5+ minutes and repeat the sanding
instructions.
f) Install a new DO membrane cap:
·
If you
need electrolyte solution, prepare a bottle of it according to the instructions
on the bottle in the Membrane Cap Kit.
·
Holding
the new cap upside-down, put in 15 drops of the solution.
·
With
the cap still upside-down, screw the cap onto the probe until it is snug—but
don’t overtighten. Some membrane
solution will spill out, indicating that the membrane cap is indeed filled.
·
Examine
the membrane; it should not be loose, wrinkled, damaged, or dirty, and there
should be no bubbles under it. If there
are problems, try rinsing, refilling, and reattaching the membrane, or discard
it and try a new one.
g) Screw the sensor guard back in.
2) Clean the conductivity cell as follows:
a) Squirt a foaming bathroom cleaner (like Dow
scrubbing bubbles) into the conductivity vents. Let sit for about 30 seconds.
b) Use the small nylon brush supplied with the
unit to dislodge any contaminants from inside the electrode chamber.
3) Rinse the probe thoroughly in clean tap
water and then purified water.
4) Make sure the sponge inside the calibration
chamber is wet, and store the probe in the chamber throughout the monitoring
session.
(see next page)
Under normal
circumstances, teams will use the YSI-85 for the entire month of monitoring
without changing the membrane or KCl solution.
To check the integrity of the DO probe during the monitoring month:
1)
Perform calibration per standard Streamkeepers protocol. The calibration number which appears on the
calibration screen in the lower-right corner, after the step in which you enter
the altitude, should be 100.0. Then let
the probe sit in tap water for an hour.
2)
Shake off the probe and put it back into the calibration chamber. Let the temperature and DO readings stabilize
(DO% within one full percentage point and temp within 0.1°C
for 30 sec.). The DO % Sat. reading
should be 100% +/- 2%. If not, the
membrane should be replaced (see above).
(Actually, replacement is not necessary, but the probe would need to be
rechecked in the field after each reading by placing it into the calibration
chamber for 15 min. and then noting the DO % Sat. reading, using the above
criterion.)
Between monitoring
sessions, the DO probe can remain in the chamber and continue to be used,
provided the integrity check is performed (see above) or the membrane is
replaced.
For long-term
storage of the meter (longer than a month):
1) Rinse the probe in a mild bleach solution
(one capful of bleach per gallon of water).
2) Holding the probe upside-down, remove the
DO membrane cap and discard the KCl solution inside. If there is no solution left under the
membrane, there may have been a problem with leakage. Note which probe had this problem and check
against the last time it was used in the field.
3) Rinse the cap and probe with distilled
water. Shake off the water.
4) Let probe and cap air dry.
5) Replace the cap, without fluid, to protect
the probe from dust.
6) Store the probe dry, outside of the
calibration chamber. Put tape over the calibration
chamber opening to keep the sponge from falling out.