MAINTENANCE
& CALIBRATION PROCEDURE:
TURBIDIMETERS
These procedures are to be performed by
Streamkeeper staff. Calibrate the
turbidimeters before at least every other monitoring session. Replace primary
and secondary (sealed) reference standards as needed so as not to exceed
expiration dates. Solutions are good
until the last day of the month listed on the expiration date.
EQUIPMENT NEEDED:
·
“Maintenance & Calibration” notebook
·
HF
Scientific Turbidimeters, models DRT-15C and DRT-15CE—these should be fully
charged before calibrating
·
Sealed
secondary field standards—each meter should have a “low” vial in the .02 NTU
range and a “high” vial in the 8-10 NTU range
·
Calibration
instructions for both models (in “Maintenance & Calibration” binder).
·
Kimwipes
& glass cleaner from field kit
·
Blue
plastic “Turbidimeter Reference Standards” box containing vials marked for
different turbidity solutions. These
vials should already have index arrows on their caps—if they do not, you will
have to index them using the instructions below. After removing any cap, be sure to replace
it on the same vial!
·
Polymer
standard solutions, or…
·
If you
are using Formazin to make calibration standards, you will need to use the
Environmental Health Lab (arrange for some counter space for ½ day) and:
·
safety
goggles
·
latex
gloves
·
turbidity
free water
·
pipettes
of different sizes from card-board “Turbidimeter Standards” box
·
volumetric
flasks and pipette suction bulbs from the Env. Health Lab
CALIBRATION PROCEDURE:
1) Warm up turbidimeters for 15 minutes before
calibrating.
2) If you are using Formazin, prepare calibration solutions, using the
Formazin dilution chart on p. 4 of the instruction manual:
a) Put the suction bulb on the appropriate
pipette.
b) Use the pipette to measure Formazin into
the appropriate volumetric flask (or 400-ml beaker).
c) Fill the rest of the flask with
turbidity-free water, to the volume stated on the chart.
d) Swirl the solution and label the flask.
3) Open the blue plastic box labeled
“Reference Standards.” Pour out the
liquid in each labeled vial, capping each vial immediately after emptying it.
4) For each labeled vial:
a) Fill ¼ full with the appropriate strength
solution, recap, agitate, and discard.
b) Refill and recap with the appropriate
strength solution.
5) If you are using Polymer standard
solutions, the bottle labeled “0.0 NTU” is actually 0.02 NTU; you’ll need to
know this for calibration.
6) Any vial you’ll be using should have a
piece of tape on top with an arrow. If
it doesn’t, you’ll have to index the
vial. This is a process whereby you
mark the orientation of the vial that yields the best light transmission
through the meter:
a) A vial can only be indexed to a single meter.
b) Set the meter to the appropriate range for
the contents of the vial.
c) Put the vial into the well.
d) Note the meter reading.
e) Make a small pencil mark on the cap at the
point nearest to the white locator pin on the rim of the well.
f) Slowly rotate the vial while observing the
meter reading, and make a new mark every time you get to a new lowest
reading. When you arrive back at your
first mark, it’s a good idea to go back around backwards, to recheck the
lowest-reading point. Once you have
confirmed that point, put a piece of tape on the cap with a small arrow
pointing to that point.
7) When all of the vials have been prepared,
calibrate each turbidimeter individually, following instructions for that
model. (For the DRT-15CE, they are
available on the Web at http://www.hfscientific.com/PDF%20-%20Manuals/DRT-15ce/Manual%20DRT-15ce.pdf.)
8) Once the machine has been calibrated, clean
the sealed secondary “low” and “high” reference vials and take readings for
them; the company guarantees that they will read within 20% of their listed
value. (If they don’t, contact the
company.) On a piece of tape on the cap
that includes a properly-oriented index arrow, write “Ref. Std. [reading on
meter] NTU / DO NOT OPEN!” Record these
readings for the secondary standards in the calibration notebook.
9) Put a new label on the control panel of the
turbidimeter, reading “Last Calibration [date][initials].”
10) After calibrating:
a) If you are using Formazin:
i) Pour the Formazin into the sink.
ii) Rinse vials, flasks, and pipettes with tap
water, then distilled water.
iii) Fill vials with distilled water, cap, and
store.
b) If you are using polymer solutions, leave these solutions in the vials to
store them.
MAINTENANCE & CALIBRATION
NOTEBOOK:
Record all calibration and maintenance
activity in the “Streamkeepers Maintenance & Calibration” binder kept in
the Streamkeepers office. In the case of
calibration, record the type of calibration solution, vendor, expiration date,
and readings on the secondary field standards.
This provides confirmation that the secondary field standards were
checked.
CALIBRATION TROUBLESHOOTING:
If
you are having trouble calibrating the instrument, you may have one of the
following problems: